The 21CFR part 11 module requires at least one user to remember the master administrator password to activate the security software.  If the master administrator password is forgotten, the following procedure should be followed.  The system will NOT remember the user detail after this procedure has been completed.

Resetting the password can achieved by this process:

  1. Copy D:\VB\Application\.cfg, where D is the CD drive, to C:\Solution11\.cfg.
  2. Replace old file (maybe backup up old one before hand) 
  3. Start Overlord, default username "paa", password "2fernhill".
  4. Message "password has expired" appears. enter new password. at least 7 characters including numbers.
  5. Resetting the password will lose current user information, but the audit trail remains. note that audit trail may appear inconsistent as it may show that the same user has been added twice (once after reset), which is normally not possible.
Click to view of download documentdownload app note
CFRTN101: Resetting 21CFR Passwords

laboratory automation - process acoustics - IT compliance (21CFR part 11 )
Copyright 2004-2012 Process Analysis & Automation Ltd.
contact | privacy policy | webpage terms | privacy policy | ISO 9001:2008

This website renders badly on Firefox v9.0.1. paa are working on this problem. It has been cause by Mozilla.

All manufacturers' trademarks are gratefully acknowledged.
paa, Overlord and automate.it are the trademark of Process Analysis & Automation Ltd.
Every effort is made to ensure the information on this website is accurate, paa cannot be held responsible for any errors.
paa and our partners reserve the right to update and change product information at any time without notice.